AFE5808AZCF
德州儀器
- Lifecycle
- Active
- Stock
- 4985 件
- Package
- NFBGA-135
- Series
- 資料獲取-類比前端(AFE)
對 AFE5808AZCF 和 AFE5818ZBV 的深入比較,為您帶來關於其規格和關鍵特性的寶貴見解。我們詳細介紹了重要因素,包括 RoHS 合規性、REACH 狀態、系列、安裝方式、封裝類型以及其他相關特性。並排展示差異,簡化了元件選擇,讓您更輕鬆地為您的特定應用選擇最佳方案。
Replacement verdict
The AFE5818ZBV and AFE5808AZCF are both 8-channel ultrasound analog front ends from Texas Instruments sharing the same core architecture, making the AFE5818ZBV a viable functional replacement candidate for the AFE5808AZCF in many ultrasound receive-chain designs. However, the two devices differ in package footprint, supply configuration, and certain performance parameters, so a direct drop-in substitution is not guaranteed. Designers should validate pinout, power sequencing, and clocking requirements against the target board before committing to the replacement.
德州儀器
德州儀器
Rows are prioritized by design impact. Highlighted values require attention during substitution.
Key electrical and mechanical differences between the two devices
| Parameter | AFE5818ZBV | AFE5808AZCF | Why it matters |
|---|---|---|---|
| Number of Channels | 16 | 8 | Channel count determines how many transducer elements can be processed simultaneously; a 16-channel device cannot directly replace an 8-channel device in a fixed 8-channel system without redesign. |
| ADC Resolution | 14 bit | 12 bit | Higher ADC resolution provides greater dynamic range and finer signal quantization, affecting image quality and small-signal detection in ultrasound receive chains. |
| Maximum ADC Sampling Rate | 65 MSPS | 65 MSPS | Sampling rate sets the maximum signal bandwidth and Nyquist limit; matching rates support comparable time-gain and beamforming timing budgets. |
| LNA Gain Range | 24 dB (fixed gain options) | 24 dB (fixed gain options) | Low-noise amplifier gain affects input-referred noise and sensitivity; comparable gain settings preserve front-end noise performance when substituting. |
| Total Power per Channel | ~150 mW/channel | ~185 mW/channel | Power per channel impacts thermal design and system power budget; a lower per-channel power device may allow denser arrays or reduced cooling requirements. |
| Package Type | NFBGA (ZBV) | NFBGA (ZCF) | Package footprint and ballout must match for board-level drop-in compatibility; differing package codes indicate different mechanical and pinout constraints. |
| Operating Temperature Range | 0°C to 70°C | 0°C to 70°C | Temperature range defines the environments where the device is guaranteed to meet specifications; matching ranges support equivalent application conditions. |
| Supply Voltage | 1.8 V / 3.3 V | 1.8 V / 3.3 V | Supply rails must be compatible with the existing power architecture; identical rails simplify substitution without regulator changes. |
| Continuous Wave (CW) Doppler Support | Yes | Yes | CW Doppler capability is required for blood-flow velocity measurement; both supporting it preserves the clinical feature set. |
| Digital Interface | LVDS | LVDS | The output interface must match the downstream FPGA/ASIC receiver; identical LVDS signaling allows reuse of the existing digital back-end. |
Send your BOM and application constraints for technical and sourcing review.
Use manufacturer datasheets as the final authority.
| Specification | AFE5808AZCF | AFE5818ZBV |
|---|---|---|
| Part Status | Active | Active |
| Number of Bits | 12, 14 | 14 |
| Number of Channels | 8 | 16 |
| Power (Watts) | 158 mW | 140 mW |
| Voltage - Supply, Analog | 1.7V ~ 1.9V, 3.15V ~ 3.6V, 5V | 1.7V ~ 1.9V, 3.15V ~ 3.6V, 5V |
| Voltage - Supply, Digital | 1.7V ~ 1.9V | 1.15V ~ 1.25V, 1.7V ~ 1.9V |
| Mounting Type | Surface Mount | Surface Mount |
The AFE5818ZBV and AFE5808AZCF are fully integrated analog front-end (AFE) devices from Texas Instruments designed for ultrasound imaging systems. Both integrate a low-noise amplifier (LNA), voltage-controlled attenuator (VCA), programmable gain amplifier (PGA), and a 12-bit or 14-bit analog-to-digital converter (ADC) with digital signal processing features.
The AFE5818ZBV is an 8-channel AFE, while the AFE5808AZCF is also an 8-channel device. Both are 8-channel analog front-ends intended for multi-channel ultrasound receive signal chains.
The AFE5818ZBV integrates a 14-bit ADC, whereas the AFE5808AZCF integrates a 12-bit ADC. The higher resolution of the AFE5818ZBV provides finer quantization for applications requiring greater dynamic range.
The AFE5818ZBV supports a maximum ADC sampling rate of 65 MSPS, while the AFE5808AZCF supports a maximum sampling rate of 65 MSPS as well. Both are suitable for high-frequency ultrasound signal acquisition.
The AFE5818ZBV is offered in a 289-ball NFBGA package, and the AFE5808AZCF is offered in a 289-ball NFBGA package. Both share the same ball-grid array footprint family, but pin assignments and thermal characteristics should be verified against the respective datasheets before any board-level substitution.
The AFE5818ZBV typically consumes approximately 1.1 W total at 65 MSPS across 8 channels, while the AFE5808AZCF typically consumes approximately 0.9 W at 65 MSPS across 8 channels. Actual power depends on operating mode, gain settings, and sampling rate.
No, they are not pin-for-pin or functionally interchangeable without design validation. Differences in ADC resolution (14-bit vs 12-bit), power consumption, and register map details require firmware and hardware verification against the respective Texas Instruments datasheets before substitution.
Both devices provide LVDS serial output interfaces for ADC data and support SPI-compatible serial control interfaces for register configuration. The exact lane count, bit depth per frame, and timing parameters differ between the 14-bit AFE5818ZBV and the 12-bit AFE5808AZCF and must be confirmed from the manufacturer datasheets.