AFE5808AZCF
德州儀器
- Lifecycle
- Active
- Stock
- 4985 件
- Package
- NFBGA-135
- Series
- 資料獲取-類比前端(AFE)
對 AFE5808AZCF 和 AFE5804ZCF 的深入比較,為您帶來關於其規格和關鍵特性的寶貴見解。我們詳細介紹了重要因素,包括 RoHS 合規性、REACH 狀態、系列、安裝方式、封裝類型以及其他相關特性。並排展示差異,簡化了元件選擇,讓您更輕鬆地為您的特定應用選擇最佳方案。
Replacement verdict
The AFE5808AZCF and AFE5804ZCF share the same Texas Instruments AFE58xx analog front-end family, ZCF 289-ball BGA package, and 12-bit LVDS output architecture, making them viable functional replacement candidates in ultrasound and medical imaging signal chains. However, the AFE5808AZCF offers eight channels versus four on the AFE5804ZCF, and its higher per-channel power and current draw require thermal and power-budget validation before substitution. Designers should confirm channel-count requirements, clock and reference configuration, and power delivery headroom prior to board-level adoption.
德州儀器
德州儀器
Rows are prioritized by design impact. Highlighted values require attention during substitution.
Key electrical and mechanical differences between the two devices
| Parameter | AFE5808AZCF | AFE5804ZCF | Why it matters |
|---|---|---|---|
| Number of Input Channels | 8 | 8 | Channel count determines how many transducer elements can be processed simultaneously; both are 8-channel devices, so system architecture compatibility is maintained. |
| ADC Resolution | 12 bit | 12 bit | ADC bit depth sets the dynamic range and quantization noise floor of the receive path; identical resolution means comparable image quality potential. |
| Maximum ADC Sampling Rate | 65 MSPS | 50 MSPS | Higher sampling rate supports wider signal bandwidth and better time resolution; the AFE5808A can digitize faster signals than the AFE5804. |
| LNA Gain Range | 0 to 24 dB (in steps) | 0 to 24 dB (in steps) | Low-noise amplifier gain range affects sensitivity and the ability to handle weak echoes; matching ranges support similar front-end tuning. |
| Programmable Gain Amplifier (PGA) Range | 0 to 40 dB | 0 to 40 dB | PGA range determines the time-gain-compensation depth profile; identical range allows reuse of the same gain control firmware. |
| Continuous Wave (CW) Doppler Support | Yes (integrated CW mixer) | Yes (integrated CW mixer) | CW Doppler capability is required for blood-flow velocity measurement; both devices include it, preserving Doppler functionality. |
| Supply Voltage | 3.3 V (analog), 1.8 V (digital) | 3.3 V (analog), 1.8 V (digital) | Supply rails must match the board power design; identical rails allow drop-in power compatibility. |
| Package | 100-pin NFBGA (ZCF) | 100-pin NFBGA (ZCF) | Package and pinout determine PCB footprint compatibility; matching ZCF packages enable direct board-level substitution. |
| Operating Temperature Range | 0°C to 70°C | 0°C to 70°C | Temperature range affects suitability for the target environment; identical commercial ranges mean equivalent environmental ratings. |
Send your BOM and application constraints for technical and sourcing review.
Use manufacturer datasheets as the final authority.
| Specification | AFE5808AZCF | AFE5804ZCF |
|---|---|---|
| Supplier | - | Rochester Electronics, LLC,Texas Instruments |
| Part Status | Active | Active |
| Number of Bits | 12, 14 | 12 |
| Number of Channels | 8 | 8 |
| Power (Watts) | 158 mW | 101 mW |
| Voltage - Supply, Analog | 1.7V ~ 1.9V, 3.15V ~ 3.6V, 5V | 3.15V ~ 3.47V, 5V |
| Voltage - Supply, Digital | 1.7V ~ 1.9V | 1.7V ~ 1.9V |
| Mounting Type | Surface Mount | Surface Mount |
The AFE5808AZCF is an 8-channel fully integrated analog front end with a maximum sampling rate of 65 MSPS and 14-bit ADC resolution, while the AFE5804ZCF is an 8-channel analog front end with a maximum sampling rate of 50 MSPS and 12-bit ADC resolution. The AFE5808A also integrates CW (continuous wave) Doppler processing, whereas the AFE5804 does not include CW Doppler mixers.
Both devices are offered in the same 100-pin TQFP (ZCF) package, but they are not drop-in replacements. Pin functions and internal signal routing differ between the two devices, so designers must verify the pinout against the respective datasheets before considering any substitution.
The AFE5808AZCF supports a maximum ADC sampling rate of 65 MSPS, while the AFE5804ZCF supports a maximum sampling rate of 50 MSPS. The higher rate of the AFE5808A enables wider bandwidth signal acquisition in ultrasound receive chains.
No. The AFE5808AZCF integrates CW Doppler mixers for continuous wave processing, while the AFE5804ZCF does not include CW Doppler functionality. Designs requiring CW Doppler must use the AFE5808A or a comparable device with integrated CW mixers.
The AFE5808AZCF provides 14-bit ADC resolution, and the AFE5804ZCF provides 12-bit ADC resolution. The higher resolution of the AFE5808A supports improved dynamic range in demanding ultrasound imaging applications.
Both the AFE5808AZCF and AFE5804ZCF are intended for medical ultrasound imaging systems, where they provide multi-channel signal conditioning, variable gain amplification, and analog-to-digital conversion for transducer receive channels.
Direct substitution is not recommended. The AFE5804ZCF has a lower maximum sampling rate (50 MSPS vs. 65 MSPS), lower ADC resolution (12-bit vs. 14-bit), and lacks integrated CW Doppler mixers. Any substitution requires full validation against the system requirements and manufacturer documentation.